EasyManuals Logo

Keithley 3730 Reference Manual

Keithley 3730
936 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #695 background imageLoading...
Page #695 background image
Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-316 3700AS-901-01 Rev. D/June 2018
Example 3
scan.create("3001:3010", "testDCV")
For this example, assume testDCV is a
previously defined user DC volts
configuration.
Clears the old scan list and creates a new
scan list with each channel (1 to 10 on
slot 3).
Each channel uses the DMM configuration
associated with testDCV.
Example 4
scan.create("")
for chan = 3001, 3010 do
scan.add("" .. chan, "testDCV")
scan.add("" .. chan, "test2wire")
end
For this example, assume testDCV is a
previously defined user DC volts
configuration, and test2wire is a
previously defined user 2-wire ohm
configuration.
This loops through the channels, adding
channels to the scan list. The first time, it
adds "testDCV" for a channel. The second
time, it adds "test2wire" for that channel.
The first parameter ("" .. chan) converts
the chan number to a string.
Clears the old scan list and creates a new
scan list.
Loops through channels 3001 to 3010.
Adds channels 3001 through 3010 to the end
of the existing scan list.
Also see
dmm.getconfig() (on page 11-181)
dmm.setconfig() (on page 11-227)
scan.add() (on page 11-309)
scan.reset() (on page 11-324)
Scanning and triggering (on page 5-23)

Table of Contents

Other manuals for Keithley 3730

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3730 and is the answer not in the manual?

Keithley 3730 Specifications

General IconGeneral
BrandKeithley
Model3730
CategoryMultimeter
LanguageEnglish

Related product manuals