7-22 Wave Functions (6221 Only) Model 6220/6221 Reference Manual
Return to Section 7 topics
Programming sine waves
1. Restore defaults with this command:
*RST
2. Configure the waveform.
Examples – The following commands configure 1kHz sine wave with an
amplitude of 10mA, 1mA offset, and phase marker off:
3. Arm and trigger the waveform, turn on output:
4. To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:FUNC SIN Select sine wave.
SOUR:WAVE:FREQ 1e3 Set frequency to 1kHz.
SOUR:WAVE:AMPL 1e-2 Set amplitude to 10mA.
SOUR:WAVE:OFFS 1e-3 Set offset to 1mA.
SOUR:WAVE:PMAR:STAT OFF Turn off phase marker.
SOUR:WAVE:DUR:TIME 5 5 second duration.
SOUR:WAVE:RANG BEST Select best fixed source range.
SOUR:WAVE:ARM ‘Arm waveform.
SOUR:WAVE:INIT Turn on output, trigger waveform.
SOUR:WAVE:ABOR Stop generating waveform.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6221 and is the answer not in the manual?

Keithley 6221 Specifications

General IconGeneral
InterfaceGPIB, RS-232
Communication InterfaceGPIB, RS-232
Current Sourcing Accuracy±(0.03% of output + 20 pA)
Current Sinking Accuracy±(0.03% of output + 20 pA)
Output Resolution100 fA
Output Accuracy±(0.03% of output + 20 pA)
Power Requirements100 V to 240 V AC, 50/60 Hz

Related product manuals