Model 6220/6221 Reference Manual Wave Functions (6221 Only) 7-25
Return to Section 7 topics
Programming arbitrary waveforms
1. Restore defaults with this command:
*RST
2. Configure the waveform.
Examples – The following commands configure 100kHz arbitrary wave
with an amplitude of 25mA, 0mA offset, phase marker off, and 20 second
duration:
3. Arm and trigger the waveform, turn on output:
4. To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:ARB:DATA -0.5, -0.25,
0, 0.3, 0.4
Define arbitrary data points in
location 0.
SOUR:WAVE:ARB:COPY 1 Copy points to location 1.
SOUR:WAVE:FUNC ARB1 Select arb wave, location 1.
SOUR:WAVE:FREQ 1e5 Set frequency to 100kHz.
SOUR:WAVE:AMPL 25e-3 Set amplitude to 25mA.
SOUR:WAVE:OFFS 0 Set offset to 0.
SOUR:WAVE:PMAR:STAT OFF Turn off phase marker.
SOUR:WAVE:DUR:TIME 20 20 second duration.
SOUR:WAVE:RANG BEST Select best fixed source range.
SOUR:WAVE:ARM ‘Arm waveform.
SOUR:WAVE:INIT Turn on output, trigger waveform.
SOUR:WAVE:ABOR Stop generating waveform.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6221 and is the answer not in the manual?

Keithley 6221 Specifications

General IconGeneral
InterfaceGPIB, RS-232
Communication InterfaceGPIB, RS-232
Current Sourcing Accuracy±(0.03% of output + 20 pA)
Current Sinking Accuracy±(0.03% of output + 20 pA)
Output Resolution100 fA
Output Accuracy±(0.03% of output + 20 pA)
Power Requirements100 V to 240 V AC, 50/60 Hz