Model 6220/6221 Reference Manual Triggering 8-11
Return to Section 8 topics
1. INIT ‘ Initiate one trigger cycle
This command is used to initiate one trigger cycle and take the unit out of the 
idle state.
2. ABOR ‘ Reset trigger system
This command resets the trigger system. If a sweep is in progress, it will be 
halted and the source value set to zero. The operate state, however, will be 
left on. The sweep is still armed, and can be started again with another INIT 
command (i.e., - no re-arming is needed).
3. ARM:SOUR <name> ‘ Select ARM layer control source.
This command selects the ARM layer event detector control source: 
IMMediate, TIMer, BUS, TLINk, NSTest, PSTest, BSTest, or MANual.
4. ARM:SIGN ‘ Bypass ARM layer control source.
This command is used to bypass the ARM layer control source and move on 
to the next layer in the trigger model (the TRIGger layer event detection in this 
case).
5. ARM:TIM <n> ‘ Set ARM layer timer interval.
This command selects the ARM layer timer interval in the range of 0 to 
99999.99 sec. with 1 msec resolution.
6. ARM:DIR <name> ‘ Set ARM source bypass.
This command enables (SOURce) or disables (ACCeptor) the ARM source 
bypass.
7. ARM:ILIN <NRf> ‘ Set ARM layer trigger input line.
This command selects the ARM layer input signal line: 1, 2, 3, 4, 5, or 6.
8. ARM:OLIN <NRf> ‘ Set ARM layer trigger output line.
This command selects the ARM layer input signal line: 1, 2, 3, 4, 5, or 6. Note 
that for the Model 6221 only, the trigger output line and phase marker 
(SOUR:WAVE:PMAR:OLIN) lines cannot be the same, or a -221 Settings 
Conflict error will occur. (See “Phase marker,” on page 7-9.).
9. ARM:OUTP <name> ‘ Select when to output ARM trigger.
This command selects when to output the ARM layer trigger. Available selec-
tions include TEXit (output when exiting trigger layer), TENTer (output when 
entering trigger layer), or NONE (disabled).
10. TRIG:SOUR <name> ‘ Select TRIG layer control source.
This command selects the TRIG layer event detector control source: 
IMMediate and TLINk.
11. TRIG:SIGN ‘ Bypass TRIG layer control source.
This command is used to bypass the TRIGger layer control source and move 
on to the next layer in the trigger model (the device action block in this case).
 
 