Model 6220/6221 Reference Manual Calibration 17-15
Return to Section 17 topics
3. Repeat step 2 for each range, as listed in Table 17-4, page 17-16, noting
the following:
Be sure to set the correct range using the RANG command.
Each range is calibrated with three values, +FS (positive full scale), 0,
and -FS (negative full scale) in that order. Use the SOUR:CURR com-
mand as shown to set the output current value for each step.
For the 2nA and 20nA ranges, use the picoammeter with connections
shown in Figure 17-2. Substitute the picoammeter reading for the DMM
reading in the command listings. Be sure to turn the output off (send
OUTP OFF) before changing connections, then turn the output back on
(OUTP ON) before completing calibration.
4. For the Model 6221 only:
a. Disconnect the DMM from the Model 6221 OUTPUT jack.
b. Send this command to calibrate the waveform generator:
CAL:PROT:WAVE:STEP1 Calibrate waveform generator
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6221 and is the answer not in the manual?

Keithley 6221 Specifications

General IconGeneral
InterfaceGPIB, RS-232
Communication InterfaceGPIB, RS-232
Current Sourcing Accuracy±(0.03% of output + 20 pA)
Current Sinking Accuracy±(0.03% of output + 20 pA)
Output Resolution100 fA
Output Accuracy±(0.03% of output + 20 pA)