EasyManuals Logo

Keithley SourceMeter 2600 Series User Manual

Keithley SourceMeter 2600 Series
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #223 background imageLoading...
Page #223 background image
Series 2600 System SourceMeters Reference Manual Source-Measure Concepts 8-17
Return to Section 8 topics 2600S-901-01 Rev. A / May 2006
Figure 8-7
I-Source operating examples
Voltage Limit
Load Line
Operating
Point
Current Source
Load Line
I-Source (I
S
)
I-Source (I
S
)
V-Meter
(V
M
)
V-Meter
(V
M
)
5V
10V
100mA
50W DUT Load Line (R)
A) Normal I-source operation
Voltage Limit
Load Line
Operating
Point
Current Source
Load Line
10V
100mA
B) I-source in compliance
50mA
200W DUT Load Line (R)
V-Meter = I
S
 R
= (100mA) (50W)
= 5V
I
S
= V
M
/ R
= 10V / 200W
= 50mA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2600 Series and is the answer not in the manual?

Keithley SourceMeter 2600 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2600 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals