EasyManuals Logo

Keithley SourceMeter 2600 Series User Manual

Keithley SourceMeter 2600 Series
594 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #237 background imageLoading...
Page #237 background image
Series 2600 System SourceMeters Reference Manual Source-Measure Concepts 8-31
Return to Section 8 topics 2600S-901-01 Rev. A / May 2006
Figure 8-15
Comparison of unguarded and guarded measurements
Insulator
I
D
DUT
Metal Mounting Plate
I
L
R
L1
R
L2
Insulator
I
M
= I
D
+ I
L
I
M
= Measured current
I
D
= DUT current
I
L
= Leakage current
IN/OUT
HI
I-Meter
IN/OUT
LO
V-Source
A. Unguarded
I
D
DUT
Metal Mounting Plate
R
L1
Insulator
I
M
= I
D
IN/OUT
HI
I-Meter
IN/OUT
LO
V-Source
x1
GUARD
0V
SourceMeter
SourceMeter
B. Guarded
Connect to earth safety ground
using #18 AWG wire or larger.
Cable Shield
Safety Shield

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2600 Series and is the answer not in the manual?

Keithley SourceMeter 2600 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2600 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals