Series 2600 System SourceMeters Reference Manual Instrument Control Library 12-35
Return to Section 12 topics 2600S-901-01 Rev. A / May 2006
Remarks Setting this attribute selects the display screen for the front panel. This attribute can be
read to determine which of the four available display screens was last selected
by the user. The user can select the screen by value or one of the enumerations.
Details See “Display screen” on Section 14.
Example Selects the source-measure and compliance limit display for SMUA:
display.screen = display.SMUA
display.sendkey
Function Sends a keycode to simulate the action of a front panel control.
Usage
display.sendkey(keycode)
Set keycode to one of the values shown below:
display.KEY_AUTO or 73 display.KEY_OUTPUTA or 88
display.KEY_CONFIG or 80 display.KEY_OUTPUTB or 96
display.KEY_DIGITSA or 87 display.KEY_RANGEDOWN or 81
display.KEY_DIGITSB or 84 display.KEY_RANGEUP or 65
display.KEY_DISPLAY or 72 display.KEY_RECALL or 85
display.KEY_ENTER or 82 display.KEY_RELA or 70
display.KEY_EXIT or 75 display.KEY_RELB or 67
display.KEY_FILTERA or 77 display.KEY_RIGHT or 103
display.KEY_FILTERB or 74 display.KEY_RUN or 71
display.KEY_LEFT or 104 display.KEY_SPEEDA or 94
display.KEY_LIMITA or 93 display.KEY_SPEEDB or 91
display.KEY_LIMITB or 90 display.KEY_SRCA or 79
display.KEY_LOAD or 95 display.KEY_SRCB or 76
display.KEY_MEASA or 86 display.KEY_STORE or 78
display.KEY_MEASB or 83 display.KEY_TRIG or 92
display.KEY_MENU or 68 display.WHEEL_ENTER or 97
display.KEY_MODEA or 69 display.WHEEL_LEFT or 107
display.KEY_MODEB or 66 display.WHEEL_RIGHT or 114
Remarks Sending this command simulates the pressing of a front panel key or wheel, or turning
the Wheel one click to the left or right.
Details See “Sending keycodes” on Section 14.
Example “Press” the RUN key:
display.sendkey(display.KEY_RUN)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2600 Series and is the answer not in the manual?

Keithley SourceMeter 2600 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2600 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals