Table 15-4
Model 2601/2602 voltage measurement accuracy limits
Model 2601/2602 source
and measure range
1
Source voltage
2
Model 2601/2602 voltage
reading limits
(1 year, 18°C–28°C)
100mV 90.000mV 89.8365 to 90.1635mV
1V 0.90000V 0.899665 to 0.900335V
6V 5.4000V 5.39819 to 5.40181V
40V 36.000V 35.9866 to 36.0134V
1. Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Table 15-5
Model 2611/2612 voltage measurement accuracy limits
Model 2611/2612 source
and measure range
1
Source voltage
2
Model 2611/2612 voltage
reading limits
(1 year, 18°C–28°C)
200mV 180.000mV 179.748 to 180.252mV
2V 1.80000V 1.79938 to 1.80062V
20V 18.0000V 17.9923 to 18,0077V
200V 180.000V 179.923 to 180.077V
1. Measure range coupled to source range when simultaneously sourcing and measuring voltage.
2. As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Series 2600 System SourceMeters Reference Manual Performance Verification 15-11
Return to Section 15 topics 2600S-901-01 Rev. A / May 2006

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2600 Series and is the answer not in the manual?

Keithley SourceMeter 2600 Series Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2600 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals