7. Operation
7.2 Starting operation
Starting and loading the device:
1. Fill the water tray with up to max. 3.0 l of processed water. Do not exceed
the upper level mark.
2. Make sure that the CO
2
/N
2
/O
2
supply system valves are open.
3. Turn the device on using the power switch.
4. Set nominal values for temperature and CO
2
/ O
2
content at the operating
panel.
5. Ventilate work space by leaving both device doors open until acoustic alarm
sounds.
6. Start device using auto-start routine.
7. Close device doors.
8. The temperature control adjusts the temperature to the set nominal value,
humidity rises.
9. When temperature and relative humidity are constant, the automatic adju-
stment of the CO
2
/O
2
measuring system is performed.
10.The ”auto-start” indicator goes off.
11.The CO
2
/O
2
control supplies the set amount of CO
2
/O
2
.
12.The device is ready for operation.
13.Load work space with cultures.
NOTE – Duration of the auto-start routine
When the device is cold and when the ambi-
ent temperature is low, the auto-start routine
may take up to 10 hours.
NOTE – Charge
To ensure sufficient air circulation and even
heating of the samples, the charge surface
within the work space should be used up to 70
% max. Voluminous objects in the work spa-
ce that dissipate heat may impair heat
distribution.