770/GS/GJ-BA-defsi-0023 125
Activating a Data Output Process
You can define the data output parameter so that output is activated either automatically
or when a print command is received. You have two options for the automatic mode:
data output can be either synchronous with the balance display or activated at defined
intervals.
Data Output by Print Command
The print command can be transmitted by a software command or by pressing a key. In
addition to an interface cable for a different device, you can connect an external
universal switch for remote control to the balance interface port (print function: see part
2). For the switch, use pins 8 and 15 of this port and a cable up to 1.5 m or 5 ft. long
(RS-232C). If data output is requested by a software command (see the section on
“Data Input Formats“), you can install a 15 m (50ft.) cable for RS 232 C.
Automatic Data Output
In the “auto print“ operating mode, the data are output to the interface port without
requiring a print command. You can choose to have data output automatically at defined
print intervals or without the stability parameter.
If you select the auto print setting, data will be transmitted immediately the moment you
turn on the balance.
Higher Data Output Rates
If you require an output rate higher than 10Hz, ask
KERN
directly for this information.