EasyManua.ls Logo

KERN EW - User Manual

KERN EW
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen-Frommern
E-Mail: info@kern-sohn.com
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.com
GB
Service Manual
Electronic Precision Balances
Page 2
KERN EW · EW-C · EG (N)
Version 2.2 01/2007
- 1 - EW_EW-C_EG_N-SH-e-0722

Questions and Answers

  • T
    thomasmichelleAug 13, 2025
    What to do if the display suddenly disappeared in operation on KERN EW?
    • J
      joshua98Aug 13, 2025
      If the display on your KERN Scales suddenly disappears during operation, examine the built-in battery, the AJDP board, and the AC adapter to identify any potential issues.
  • K
    kellyloriAug 17, 2025
    Why does the display not get settled down on KERN EW Scales?
    • P
      Paula PetersonAug 17, 2025
      If the display of your KERN Scales does not settle down, doesn't repeat correctly, or the zero point drifts, inspect the stopper parts, check for any interference with the weighing pan, look for any foreign substances, and examine the tuning-fork sensor. Also, consider the impact of wind or oscillation on the scales.
  • M
    Michael QuinnAug 21, 2025
    How to check span is out of specified range for KERN Scales?
    • C
      Christine LewisAug 21, 2025
      If the span is out of the specified range on your KERN Scales, examine the tuning-fork sensor, the AJDP board, the address data, and the coefficient memories.
  • S
    Seth FryeAug 25, 2025
    How to check linearity is out of specified range for KERN EW?
    • C
      Christopher DownsAug 25, 2025
      If the linearity is out of the specified range on your KERN Scales, check the tuning-fork sensor, address data, coefficient memories, and the calibration weight.
  • N
    Nicole WhiteAug 28, 2025
    How to fix no display lights on KERN EW Scales?
    • T
      Tina RodgersAug 28, 2025
      If the display lights are not turning on for KERN Scales, examine the AJDP board, AC adapter, connection cords, or battery.
  • B
    Bryan GarciaSep 1, 2025
    What to do if o-Err appears with a net load less than specified capacity on KERN EW?
    • T
      Timothy WilsonSep 1, 2025
      If an 'o-Err' appears on your KERN Scales with a net load less than the specified capacity (F.S. +9d), check the applied gross weight, coefficient memories, address data, or calibration weight.
  • H
    Harry BowmanSep 5, 2025
    What to do if u-Err or o-Err appears self test of segments on KERN Scales?
    • A
      Alec BrownSep 5, 2025
      If 'u-Err' or 'o-Err' appears during the self-test of segments on your KERN Scales, check the weighing pan, tuning-fork sensor, AJDP board, address data, or coefficient memories.
  • H
    hwhiteSep 8, 2025
    How to resolve corner error is too much on KERN EW?
    • M
      Megan KellySep 8, 2025
      If the corner error is too much on your KERN Scales, inspect the mechanism unit and the pan base.

Summary

1 TOTAL VIEW

1.1 Total View Of EW Type - 220-12000 g

Overview of EW type models with capacities up to 12000g.

1.2 Total View Of EG Type - 220-620 g, 2200 g, 4200 g

Overview of EG type models with capacities up to 4200g.

2 ELECTRONIC CONSTRUCTION

2.1 Block Diagram

Visual representation of electronic components and connections.

2.2 Whole Wiring

Complete wiring schematic for the electronic system.

3 TROUBLESHOOTING

3.1 Troubleshooting Procedure

Systematic steps for diagnosing and resolving issues.

3.2 Trouble shooting Table

Lists common symptoms, causes, and remedies.

4 ADJUSTMENTS AND SETTINGS

4.1 Span Calibration (CAL)

Procedure for calibrating the scale's span.

4.3 Corner Error Adjustment

Steps to adjust for corner errors on the weighing pan.

4.5 Resetting Address Data (Coefficients)

Procedure for resetting stored address data or coefficients.

4.7 Linearity Adjustment

Steps to perform linearity adjustment on the scale.

5 PARTS REPLACEMENTS

5.1 How to remove the case

Instructions on how to detach the scale's outer casing.

5.3 Sequence Of Mechanism Unit Replacement

Overview of the procedure for replacing the mechanism unit.

5.6 Sequence Of The AJDP Circuit Board Replacement

Overview of the procedure for replacing the AJDP circuit board.

6 INSTALLATION OPTIONS

6.1 Installation Of Battery Option

Instructions for installing an optional battery.

7 PARTS LIST

7.1 EW 220-1200

Parts list for the EW 220-1200 model.

7.2 EW 2200-12 K

Parts list for the EW 2200-12K model.

7.3 EW-C; EW 120-4 NM

Parts list for EW-C and EW 120-4 NM models.

KERN EW Specifications

General IconGeneral
DisplayLCD
Minimum part weight at piece counting (Laboratory scale)0.1 g
Weighing plate materialstainless steel
Measuring systemStrain gauge
InterfaceRS 232
CalibrationExternal calibration
Power SupplyAC adapter or 4xAA batteries
Unitsg lb oz