EasyManua.ls Logo

KERN EW

KERN EW
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Adjustments and Settings
4.1 Span Calibration (CAL)
4.1.1 EW and EW-C Type
1. It is recommended to use a calibration weight of better accuracy than the
scale.
2. The calibration is available with ½ of F.S. Nevertheless, it is recommend to
use weight closer to F.S. for accurate calibration.
3. Error messages:
o-Err
: The calibration weight is over the full capacity.
I-Err
: The calibration weight is less than ½ of the capacity.
2-Err
: The data error exceeds 1%. Or perhaps the scale may be defective.
- 12 - EW_EW-C_EG_N-SH-e-0722

Related product manuals