5.5 Sequence Of Tuning-Fork Sensor Replacement
5.5.1 How to uninstall Tuning-fork assy
(1) Remove the mechanical unit from the chassis.
(2) Replace the tuning-fork sensor.
(3) Insert shim at stopper.
(4) While holding guide link down to under direction, tighten screws.
(Need the torque 25kgfcm)
- 36 - EW_EW-C_EG_N-SH-e-0722

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EW and is the answer not in the manual?

KERN EW Specifications

General IconGeneral
DisplayLCD
Minimum part weight at piece counting (Laboratory scale)0.1 g
Weighing plate materialstainless steel
Measuring systemStrain gauge
InterfaceRS 232
CalibrationExternal calibration
Power SupplyAC adapter or 4xAA batteries
Unitsg lb oz