4.2 USING “2006” TO ENTER THE SERVICE PARAMETERS
After the jumper has been placed in the position shown, apply power to the scale.
The display will ask for a code number, “Pn” on the Weight Display.
Enter the number 2006 when “Pn” is displayed and then press [Tare].
The displays will show the first parameter “F1” “CAL”. To scroll through other
parameters press the key.
Press the [Tare] key to enter a parameter.
To exit a parameter, press the key.
The Weight Display window will show the parameter number and the Piece Weight
window will show the word describing the function.
When a parameter is entered by pressing the [Tare] key, the displays will guide you
through the parameter selected and the options available.
The parameters available are:
F1 CAL To enter the Calibration
F2 rES Resolution selection, not valid for CPB-M units
F3 CAPA Select Capacity
F4 init ZEro Initial Zero Range
F5 rE ZEro Re-Zero Range
F6 SCSivE tArE Successive Tare Enable
F7 Ad Count Display the A/D counts
F8 ZEro Set zero method.
F9 Pin Set new PIN number
F10 LVd ModE Set low voltage detect mode
CPB-SH-e-0721 11