EasyManua.ls Logo

KERN FKB A Series - Animal Weighing Function

KERN FKB A Series
37 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FKB_A-BA-e-1615
29
Animal
weighing
function
0.0
g
Vn�
T
ANL
The animal weighing function can be applied for busy weighing.
During a defined period the average value of the weighing results
is formed.
The more unquiet the weighed item, the longer the
period should
be selected.
In weighing mode keep the PRINT key pressed until [Unit]
appears.
Press the MODE button several times until „ANL“ is displayed.
Acknowledge using SET key, the current setting is displayed.
Select the desired settings by pressing the MODE key
ANL 3
Period 3 seconds
ANL 5
Period 5 seconds
ANL 10
Period 10 seconds
ANL 15
Period 15 seconds
ANL off
Animal weighing not active
Use the SET key to confirm selection.
Put the weighing good (animal) on the weighing plate and press
the SET button. In the display runs a „Countdown“.
The average value of the weighing results is displayed and
remains displayed on the screen.
Use the SET key to change between animal weighing and
normal weighing.
Press the SET key to restart the weighing cycle for animal
weighing.

Related product manuals