Calibration procedure:
If the calibration menu
is brought up after the password has been
entered, the verification is lost, as the value of the integrated counter
(<Con x>) is increased by ‘1’.
Balance indication Operation
Switch the balance off.
Press the ON/OFF button and hold it down.
When the ON/OFF is held down and Con
x value is
displayed, press TARE
3 times. Then release the
ON/OFF button.
A "0000" password request will be displayed.
Enter "9810" password.
When you press TARE
, the numerical value will be
increased; when you press UNIT
, the decimal position
is changed.
To confirm the entered data, press ON/OFF button
The integrated counter value will be displayed.
Press TARE to confirm, "0
0
" will be displayed.
(example)
Put on a calibration weight. Wait until the stabilisation
indicator is displayed, press TARE to confirm.
"CAL" will be displayed.
The balance will switch to the weighing mode.
Switch the balance off and then on again.
Con value will be increased by "1".
Th
is means the calibration has been completed
successfully.