28 FOB-BA-e-1821
Models FOB-LM:
• For balances which need to be verified, the calibration function is blocked.
• Following each calibration, the integrated counter value (<Con
increased by one.
•
If the service menu is brought up after the password has been entered, the
verification is lost, as the value of the integrated counter (<Con x>) is
increased by ‘1’.
Balance preparation before calibration:
Switch the balance off.
Press the ON/OFF button and hold it down.
When the ON/OFF is held down and Con x
value is displayed, press TARE 3 times. Then
release the ON/OFF button.
A "0000" password request will be displayed.
Enter "9810" password:
When you press TARE
will be increased; when you press UNIT
decimal position is changed.
To confirm the entered data, press ON/OFF
button
"SEt” and then "Unit” will be displayed.
Holding UNIT
down, scroll the menu forwards,
until "GrAu" is displayed.
Press TARE.
Holding UNIT
down, select "OFF" and press
TARE to confirm.
Holding UNIT down, select "End".
Press TARE
to confirm; the balance will switch
to the weighing mode automatically.