572/573/KB/DS/FKB-BA-defsi-0651
33
6.2 Unpacking
Carefully remove the balance from its packaging, remove the plastic wrapping and position the balance in its
intended working location.
6.2.1 Installation
Install the balance in such a fashion that the weighing plate is absolutely horizonta
6.3 Mains supply
Electric power supply is by means of the external mains supply circuit. The printed voltage level must comply
with the local voltage.
Only use original KERN mains supply circuits. The use of other makes is subject to approval by Kern.
6.4 Connecting peripheral equipment
The balance must be disconnected from the mains before connecting or disconnecting additional equipment
(printer, PC) to or from the data interface.
Only use KERN accessories and peripheral equipment with your balance. These have been ideally coordina-
ted to your balance.
6.5 Initial start-up
A warm-up time of 15 minutes stabilises the measured values after switching on.
The accuracy of the balance depends on the local acceleration of the fall.
Please be sure to observe the information in the chapter on adjusting.
6.6 Adjustment
As the acceleration value due to gravity is not the same at every location on earth, each balance must be
coordinated – in compliance with the underlying physical weighing principle - to the existing acceleration
due to gravity at its place of location ( only if the balance has not already been adjusted to the location in the
factory). This adjustment process must be carried out during the initial start-up, after change in location and
variation of surrounding temperature. It is also recommendable to adjust the balance periodically during
weighing operation in order to obtain exact measured values.
6.7 Adjusting
Using a precision weight, the accuracy of the balance can be checked at any time and adjusted.
Warning:
With verified balances, this precision adjustment is not possible
Adjustment procedure:
Check that the surrounding conditions are stable.
A short warm-up time of about 15 minutes is recommended for stabilisation.
Question and Answer IconNeed help?

Do you have a question about the KERN KB Series and is the answer not in the manual?

KERN KB Series Specifications

General IconGeneral
CalibrationExternal calibration
Weighing Unitsg, kg
DisplayLCD
InterfaceRS-232
Linearity±1 digit

Summary

Technical Data

Fundamental Safety Information

Operating Instructions and Staff Training Requirements

Emphasizes reading instructions and using only trained staff for operation.

Unpacking, Installation, and Commissioning

Installation and Initial Setup Procedures

Covers place of installation, unpacking, mains supply, peripherals, and initial start-up.

Balance Adjustment and Verification

Details the process of adjusting the balance and information on official verification.

Operation and Controls

Detailed Weighing and Operational Functions

Comprehensive guide to various weighing modes like Tare, Count, Percent, Formula, Tolerance.

Interface and Data Transfer

Maintenance, Troubleshooting, and Disposal

Troubleshooting Common Issues

Lists common problems and their potential causes and solutions.

Related product manuals