PFB-BA-e-1630 23
Switch off Sale mode: Setting „SALE n“ = 6000 e or 60000 d
The totalizing function is available
The weighing units gram and carat are available
Start balance by pressing . While the
balance carries out a self test, press until
„F1 Unt“ appears
Press repeatedly untilSALE n“ is
displayed
Press adjustment switch on the lower side of the balance!
(when „SALE n“ is not displayed, press )
Confirm setting by
The balance changes automatically into weighing mode
and thus will be adjusted for 6000 e or 60000 d.
Connect the second display to the balance via the delivered connecting cable
Now the balance can be controlled via the second display:
Press to switch on or off only the second display
Reset balance and second display to zero by
Tare balance and second display to zero by