R
rjohnsonSep 23, 2025
What to do with the input value for the measurement incertainty exceeds maximum setting range on KERN PES 620-3M?
- JJulie BrowningSep 23, 2025
Use adjustment weights with poor deviation for your KERN Scales.
What to do with the input value for the measurement incertainty exceeds maximum setting range on KERN PES 620-3M?
Use adjustment weights with poor deviation for your KERN Scales.
Why is the weighing result wrong on my KERN PES 620-3M?
If the weighing result on your KERN Scales is incorrect, ensure the display is at zero. If not, zero the display. Also, the balance may need re-adjusting. Make sure the balance is placed on a level surface. Extreme temperature fluctuations or electromagnetic fields could also be the cause, so control the ambient temperature and consider moving the scale to a different location or switching off any interfering devices.
What to do if the displayed weight is permanently changing on my KERN PES 620-3M Scales?
If the displayed weight on your KERN Scales is constantly changing, it might be due to a draught or air movement; try to eliminate these. Table or floor vibrations can also cause this issue, so stabilize the surface. Ensure the weighing plate isn't touching other objects. Electromagnetic fields or static charging could also be the reason; consider relocating the scale or switching off any nearby devices causing interference.
How to resolve maximum weighing range exceeded error on KERN PES 620-3M?
If you've exceeded the maximum weighing range on your KERN Scales, split the sample and weigh each portion individually. Alternatively, use a lighter tare weight.
What to do if there is a faulty end of the internal adjustment on KERN Scales?
Perform internal adjustment again on your KERN Scales.
What to do if weight value is too low in piece counting mode on KERN Scales?
If the weight value of the sample is too low when setting the reference weight in piece counting mode on your KERN Scales, use samples or reference weights with a higher weight value. Consider the lowest piece weight and minimum load.
What to do if KERN PES 620-3M Scales shows maximum weighing range exceeded?
If your KERN Scales show that the maximum weighing range has been exceeded, you can: * Split the sample and weigh each part individually. * Use a lighter tare weight.
What does it mean when KERN PES 620-3M show negative load below minimum weighing range?
If your KERN Scales display a negative load below the minimum weighing range, check the following: * Ensure the weighing plate or weighing plate carrier is correctly adjusted. * Verify that the balance is not touching any other objects.
What causes a negative load below the minimum weighing range on KERN PES 620-3M Scales?
Check whether the balance is touching other objects on your KERN Scales.
What to do if the weighing plate is loaded during the internal adjustment of KERN PES 620-3M?
Unload the weighing plate and repeat the internal adjustment on your KERN Scales.