17 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory).
This adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
• Observe stable environmental conditions. A warm up time (see
chapter 1) is required for stabilization.
• Ensure that there are no objects on the weighing plate.
• Avoid vibration and air draught.
• Always carry out adjustment with the standard weighing plate in
place.
• The adjustment record is printed if an optional printer is connected
and the GLP function has been activated.
17.1 Adjustment with internal weight
• This function is only available for the following weighing system: PEJ
• Cancel the process by pressing the [PRINT] key
Select <7. CA. 1> in the menu (navigation
in the menu: see chapter 8.3)
The balance returns to weighing mode
Press the [CAL] key to start internal
adjustment.