21 Instant help for troubleshooting
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
The weight display does not glow
• The balance is not switched on
• The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
• Power supply interrupted.
The displayed weight is permanently changing
• Draught/air movement
• Table/floor vibrations
• Weighing plate has contact with other
objects
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing result is obviously wrong
• The display of the balance is not at zero
• Adjustment is no longer correct
• The balance is on an uneven surface
• Great fluctuations in temperature
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing result is wrong after the
adjustment
• Adjustment was not carried out under stable
ambient conditions.
• Differences in weight between the
adjustment weight and the weight used for
testing
The display does not change when the M
symbol flashes
• Draught/air movement
• Table/floor vibrations
• Weighing plate has contact with other
objects
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
Should other error messages occur, switch balance off and then on again.