Fault
Possible cause
The weight display does not glow
The balance is not switched on
The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
Power supply interrupted.
The displayed weight is permanently changing
Draught/air movement
Table/floor vibrations
Weighing plate has contact with other
objects
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing result is obviously wrong
The display of the balance is not at zero
Adjustment is no longer correct
The balance is on an uneven surface
Great fluctuations in temperature
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
The weighing result is wrong after the
adjustment
Adjustment was not carried out under stable
ambient conditions.
Differences in weight between the
adjustment weight and the weight used for
testing
The display does not change when the M
symbol flashes
Draught/air movement
Table/floor vibrations
Weighing plate has contact with other
objects
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)