EasyManua.ls Logo

KERN PES 620-3M

KERN PES 620-3M
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Pay attention to upper and lower case letters when entering data
Wait for the balance to respond between two entries
Input format:
1
2
3
4
C1
C2
CR
LF
Example of permanent output:
Input: O0
C1
C2
Code (C1)
Code (C2)
Significance
T
Sp
54H
20H
Set to zero/taring
O
0
4FH
30H
End output
O
1
4FH
31H
Permanent output
O
2
4FH
32H
Continuous output only for stable values
(interruption of output for unstable
values).
O
3
4FH
33H
Press [PRINT] key for one-time output
O
4
4FH
34H
Automatic output when weighing plate is
loaded again and value is stable
O
5
4FH
35H
One-time output whenever value is stable
(no output for unstable values)
O
6
4FH
36H
Continuous output for unstable values
(interruption of output when value is
stable stable value is output once)
O
7
4FH
37H
Press [PRINT] key for one-time output at
stable values (no output at unstable
values)
O
8
4FH
38H
Single output
O
9
4FH
39H
One-time output at stable value
O
A
4FH
41H
Output in any preset time interval
O
B
4FH
42H
Output in any preadjusted time interval
when value is stable (interruption of the
output at instable values)

Table of Contents