TPES-B_TPEJ-B-BA-e-2319 64
Press the [S] key
An acoustic signal sounds and the
balance returns into weighing mode
16.2 Measurement incertainty of the external adjustment weight
The measurement incertainty indicates the deviation of the display from the external
adjustment weight. By entering the measuring incertainty, this deviation from the
adjustment or the adjustment test can be taken into account with an external
adjustment weight. By that way a more exact adjustment can be possible.
Measurement incertainty = Displayed weight - Nominal value
16.2.1 Enter measurement incertainty
In the enhanced menu navigate to <2.
o.M.P.> and select the setting
Enter measurement incertainty
Measurement incertainty entered:
Select <2.o.M.P. 1>
The last value stored for the measurement
incertainty is shown in mg and flashes
Press the [TARE/ZERO] key
• These functions are only available for the PES weighing system.
• If more than one adjustment weight is used, the deviations must be
totalized and entered as total measurement incertainty
• The measurement incertainty must not be more than +/- 100 mg.
Otherwise the error message <r-Err> appears.