TPES-B_TPEJ-B-BA-e-2319 48
13.9 Setting the tolerance values
• Tolerance values can only be saved when the balance is in the
measurement mode display
• Set the balance to zero (see chap. 9.2) or tare (see chapter 9.3) the
balance before saving tolerance values.
For the distinction method setting with absolute values <24. tYP. 1> (see
chap. 13.5)
Actual value setting method:
When the balance is in the measuring
mode, hold down the [S] key for about 2
seconds.
When <L. SEt> or <L1 SEt> is displayed,
release the [S] key
Last stored value for the lower tolerance
limit is appears on the display and flashes
(in this example: Weight value)