✓ Automatic calibration (applicable to KN-5000D):
Click “Irradiation calibration”, input “Intensity setting password”
(123456), following the instructions of the equipment to put the
treatment handle into the handle holder correctly. Press the “ON”
button on the handle, the equipment will enter the automatic calibration
procedure. After the calibration is finished, the equipment will show the
calibration result.
Warning!
⚫ The factory setting of irradiation intensity values are shown in
Appendix A. The irradiation intensity value should be
corrected according to the measurement results. Please do
not modify it arbitrarily to avoid inaccurate dose or time.