44
DisplayKey(s) used Operating instructions
With the value ‘0.000’ flashing, open the heater lid and remove
the weight.
Closing the heater lid and pressing the
key causes a string
of hyphens (i.e., ‘- - - -’) to be displayed, 0-point calibration to
come to an end, and then an ‘END’ message to be displayed.
* If the unit is connected to a printer, then a calibration record will
be automatically printed out.
After several seconds, the display will return to the regular weight
display.
Step
6
7