53
DisplayOperating instructions
Completion of measurement
Measuring will be completed when the conditions specified for automatic halting
mode have been met. When measuring is completed, the bias will be displayed
in the bias display area.
* If no predicted measurement is displayed even after 30 minutes has elapsed
from the time when measuring was initiated, it is probable that it is impossible
to obtain a predicted measurement.
[Measurement results]
Predicted measurement
Automatic halting mode measurement
Bias + 0.60
= 16.40 (Automatic halting mode value) – 15.80 (Predicted measurement)
Without changing any of the conditions used in steps 1 through 5, repeat
steps 6 through 8 to measure again 5 times or more and calculate the
average bias.
The average bias may then be used to specify the bias to be used in predictive
measuring mode.
* If the bias varies greatly from one time to the next when performing repeated
measurements, then it is probably safe to assume that a predicted
measurement cannot be obtained. If this happens you should measure in
automatic halting mode or in some other mode instead of predictive.
If you wish to use the calculated bias to evaluate predicted measurements,
follow the procedure described in “12-2-2 Evaluating predicted measurements”
on p. 54; if you wish to use the calculated bias to actually obtain predicted
measurements, follow the procedure described in “12-2-3 Obtaining predicted
measurements” on p. 55.
Step
8
9
0

Table of Contents