63
8-
5
Stepped mode format
“_Mode_:_Step” + delimiter
tab + “Temp (C)” + tab + “Time (min.)” + delimiter
The following output then appears the same number of times as the number of specified
steps:
“_Step” + “X” (1-byte step number) + tab + “XXX” (3-byte temperature setting) + tab +
“XXX” (3-byte drying time setting) + delimiter
8-
6
Comparative mode format
“_Mode_:_Compare” + delimiter
“_Setting Temp._:_” + “XXX” (3-byte temperature setting) + “C” + delimiter
“_Auto Stop Cond._:_” + “X.XX” (4-byte automatic halting conditions setting) + “%” +
delimiter
“_Pred. Tol._:_” + “X.X” (Predicted value convergence range) + delimiter
8-
7
Predictive mode format
“_Mode_:_Predict” + delimiter
”_Setting Temp._:_” + “XXX” (3-byte temperature setting) + “C” + delimiter
“_Pred. Tol._:_” + “X.X” (Predicted value convergence range) + delimiter
9 “_Bias_:_” (+“-”) + “X.XX” + “%” + delimiter
Measurement output format
tab + “Time (min.)” + tab + “Temp. (C)” + tab + “Mass (g)” + tab + “Moist (%)” + delimiter
Intermediate measurement output format
tab + “XXX.X” (5-byte measuring time) + tab + “XXX” (3-byte of thermistor temperature) + tab +
“XXX.XXXX” (8-byte sample weight) + tab + Moisture content “XXX.XX” (6-byte moisture content)
+ delimiter
Final measurement output format
“*” + tab + “XXX.X” (5-byte measuring time) + tab + “XXX” (3-byte of thermistor temperature) + tab
+ “XXX.XXXX” (8-byte sample weight) + tab + Moisture content “XXX.XX” (6-byte moisture con-
tent) + delimiter
Sample output
Maker
Model
Serial number
Device ID
Sample code
Date & time of measurement
Measuring conditions storage area
Measurement standard
Measurement mode
Drying temperature
Automatic halting conditions
Bias
Elapsed measuring time
Changes in drying temperature
Changes in weight
Measured value (%)

Table of Contents