10 Keysight 34921A-34925A User’s Guide
Measurement Functions for the Multiplexer Modules
The MUX modules support the DMM measurement functions shown in the 
following table.
[a] Direct current measurements are allowed on channels 41 through 44 only (for all other channels, external shunts are required). 
[b] Optional 34921T Terminal Block is required for thermocouple measurements with built-in internal reference junction.
[c] A fixed or external reference junction temperature is required for thermocouple measurement with this module.
[d] Impact of higher offset voltage specification (< 50 μV) must be taken into consideration.
[e] 1 kΩ or higher range used unless 100Ω series resistors are bypassed on module.
[f] 10 kΩ or higher range used for loads over approximately 300Ω due to series resistance of FET channels.
Measurement Function(s)
34921A
40-ch
Armature
Mux
34922A
70-ch
 Armature
Mux
34923A
40-ch 
 Reed   Mux
(2-Wire)
 34923A
80-ch   
 Reed  Mux
(1-Wire)
34924A
70-ch 
Reed   Mux)
34925A
 40-ch 
FET Mux
(2-Wire)
34925A
80-ch 
FET Mux
 (1-Wire)
 Voltage, AC/DC Yes Yes Yes Yes Yes Yes Yes
 Current, AC/DC
Yes
[a]
No No No No No No
 Frequency/Period Yes Yes Yes Yes Yes Yes Yes
 Ohms 2-Wire  Yes Yes
Yes
[e]
Yes
[e]
Yes
[e]
Yes
[f]
Yes
[f]
 Ohms 4-Wire  Yes Yes
Yes
[e]
No
Yes
[e]
Yes
[f]
No
 Thermocouple 
Yes
[b]
Yes
[c]
Yes
[c],[d]
Yes
[c],[d]
Yes
[c],[d]
Yes
[c]
Yes
[c]
 RTD 2-Wire Yes Yes
Yes
[e]
Yes
[e]
Yes
[e]
No No
 RTD 4-Wire Yes Yes
Yes
[e]
No
Yes
[e]
Yes
[f]
No
 Thermistor  Yes Yes
Yes
[e]
Yes
[e]
Yes
[e]
No No