54 Keysight 34921A-34925A User’s Guide
This drawing shows two independent 20-channel 2-wire MUXes. To change
configuration modes, use the SYSTem:MODule:WIRE:MODE command.
34925A D-Sub Connectors for Two- or Four-Wire Mode
026
027
028
029
030
031
032
033
034
035
036
037
038
039
040
021
022
023
024
025
001
002
003
004
005
006
007
008
009
010
011
012
013
014
015
016
017
018
019
020
H
L
H
L
H
L
H
L
ABus2
DMM
(SENS)
ABus3
ABus4
923
913 914
924922
912911
921
ABus1
DMM
(MEAS)
H
L
H
L
H
L
H
L
LH
LH
LHLHLHLH
LHLHLHLH
Bank 1
Bank 2
COM 1
COM 2
Analog Buses
Overvoltage Protection
(each channel)
100Ω
H
L
51.1Ω
11Ω 11Ω
100Ω
Current-Limiting Circuitry
100Ω
NOTE: The three-digit number assigned to
each switch represents the channel number.
NOTE:
Bank relays: FET non-latching
Analog Bus relays: Armature non-latching

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 34921A and is the answer not in the manual?

Keysight Technologies 34921A Specifications

General IconGeneral
BrandKeysight Technologies
Model34921A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals