Calibration Procedures 4
Keysight 34970A/34972A Service Guide 115
Tests 2 - 3:
Make the connections to the 34908A as shown in the diagram below. Be sure to
route your wiring for proper strain relief and install the module cover. Install the
34908A in slot 200 of the 34970A/34972A. Open all channels on the module by
performing a Factory Reset (press
and select “Recall State”; press
again and select “Factory Reset”). Configure Channel 01 as follows: DC volts, 10
volt range, and 5½ digits.
Connections for 34908A Verification Tests 2 and 3
Enable reading monitoring by pressing on the selected channel (or use the
ROUTe:MON command). Record the 4-wire ohms measurements from the external
DMM in the following table.