EasyManuals Logo

Keysight Technologies E3648A User Manual

Keysight Technologies E3648A
255 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
3 Remote Interface Reference
130 E364xA User’s and Service Guide
Interface Configuration Commands
Refer to also “Configuring the Remote Interface” on page 66 for more information.
SYSTem:INTerface {GPIB|RS232}
Select the remote interface. Only one interface can be enabled at a time. The
GPIB interface is selected when the power supply is shipped from the factory.
SYSTem:LOCal
This command places the power supply in the local mode during RS-232
operation. All keys on the front panel are fully functional.
SYSTem:REMote
Place the power supply in the remote mode for RS-232 operation. All keys on
the front panel, except the Local key, are disabled while in the remote mode.
SYSTem:RWLock
Place the power supply in the remote mode for RS-232 operation. This
command is the same as the SYST:REM command except that all keys on the
front panel are disabled, including the Local key.
<Ctrl-C>
Clear the operation in progress over the RS-232 interface and discards any
pending output data. This is equivalent to the IEEE-488 device clear action
over the GPIB interface.
NOTE
It is very important that you send the SYST:REM command to place the power
supply in the remote mode. Sending or receiving data over the RS-232 interface
when not configured for remote operation can cause unpredictable results.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E3648A and is the answer not in the manual?

Keysight Technologies E3648A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE3648A
CategoryPower Supply
LanguageEnglish

Related product manuals