Operation and Features 2
E364xA User’s and Service Guide 73
4 Clear the overvoltage cond ition and exit this menu.
The annunciator for the OVP operation will not flash any more. The output
will return to the meter mode.
– By adjusting the OVP trip level
1 Raise the OVP trip level higher than the level tripped
2 Select the OVP Clear mode by turning the knob.
3 Clear the overvoltage condition and exit the menu.
The annunciator for the OVP operation will not flash any more. The output
will return to the meter mode.
Remote interface operation
DONE
OVP ON
OVP CLEAR
DONE
VOLT:PROT {<voltage>|MIN|MAX} Set the OVP level.
VOLT:PROT:STAT {OFF|ON) Disable or enable the OVP circuit.
VOLT:PROT:CLE Clear the tripped OVP circuit.