Installation Note N9010-90067 25
Option EXM, External Mixing Upgrade Kit
Verify the License Installation
1. Before the licenses will be recognized, the XSA application must be restarted. Press System,
Exit Program. An Exit Analyzer dialog box will appear; press OK to confirm the exit.
2. Double-click on the LaunchXSA icon on the Windows desktop. Wait for the XSA application to
finish starting (the analyzer should be sweeping).
3. Press System, Show System on the analyzer to display a list of all displayed options. You
should see the following option listed:
N9010B-EXMExternal Mixing
Verify Optional Functionality
1. Press Mode/Meas, Spectrum Analyzer, Swept SA, OK.
2. Press Input/Output. Press Select Input.
3. Verify that “External Mixer” appears below the “RF” key.