EasyManuals Logo

Keysight Technologies EXA N9010B User Manual

Keysight Technologies EXA N9010B
27 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #6 background imageLoading...
Page #6 background image
6 Installation Note N9010-90067
Option EXM, External Mixing Upgrade Kit
Installation Procedure
1. Connect a power cord to the analyzer and turn on the analyzer.
2. After the analyzer has completed turning on, press System, Show System. Make note of the
following information from the Show System screen:
Product Number: _____________
Serial Number: ______________
Instrument S/W Revision: ______________
3. Check for the presence of one of the following frequency range options listed below in the
Show System. Put a check mark or “X” after the frequency range option listed below that
appears in the Show System menu.
N9010A-503_____
N9010A-507_____
N9010A-513_____
N9010A-526_____
N9010A-532_____
N9010A-544_____
4. If the analyzer is equipped with any of the following frequency range options (refer to the data
in step 3 above):
N9010A-503
N9010A-507
N9010A-513
N9010A-526
do not proceed with the installation of this upgrade kit.
An RF or microwave EXA is not compatible with Option EXM.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies EXA N9010B and is the answer not in the manual?

Keysight Technologies EXA N9010B Specifications

General IconGeneral
BrandKeysight Technologies
ModelEXA N9010B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals