68 Keysight FieldFox Handheld Analyzers Service Guide
Theory of Operation
Measurement Group
5-
System Monitor (SM) Microcontroller
This is the heart of the analyzer. It is active whenever there is DC power
available. The microcontroller has ultra-low power dissipation. It is sometimes
referred to as the “MSP 430”.
The following is a list of some of the functions performed by the SM.
It monitors the On/Off button on the front panel and responds
appropriately.
It controls the on/off functions of the various power supplies in the analyzer.
It drives the On/Off button LED.
It reads battery pack information via the Smart Batt Bus.
It communicates with the battery charger IC via the Smart Batt Bus.
It interfaces to the Battery Charger and Manager circuit.
It interfaces to the Serial Bus for communication with the temperature
sensors and ID EPROM’s on the printed circuit boards.
It communicates to the A6 SOM board CPU via a serial port, UART.
It stores the instrument model number and serial number.
USB Hub
This circuitry expands the number of USB ports available to the FieldFox
firmware.
Power Management and Power Supplies
Seven different analog supply voltages are generated from the Unreg DC. All of
these are used on the A4 RF board and some are used on the A5 System board.
Some sections of the circuitry on the A4 RF board may be switched to low
power modes to conserve power and minimize heat. This is accomplished
through load switching. See “Power Management Circuitry” on page 66 for
more information.
The System Monitor controls the on/off functions of the analog power supplies.
All are on or off together except for power sequencing.
This circuitry also monitors the Ext DC input and, if below 10.8 volts or above
21 volts, it disconnects this input from the Unreg DC.
FPGA
This FPGA:
provides the interface between the A6 SOM board and the ADC,
provides miscellaneous interfaces for circuitry on the A4 RF board, and
communicates with the CPU via the USB interface.

Table of Contents

Other manuals for Keysight Technologies FieldFox N9938A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies FieldFox N9938A and is the answer not in the manual?

Keysight Technologies FieldFox N9938A Specifications

General IconGeneral
BrandKeysight Technologies
ModelFieldFox N9938A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals