102 Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
6 Testing Performance
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1V/div ±3.0V ±2 %
2V/div ±6.0V ±2 %
5 V/div ±15.0 V ±2 %
Table 10 DC Gain Accuracy Test (1 M) (continued)
Vertical
Scale
Ext. Supply
Setting
V
DMM+
V
Scope+
V
DMM-
V
Scope-
Calculated
DC Gain
Error
DC Gain
Error Test
Limits
Table 11 Analog Bandwidth—Maximum Frequency Test
Vertical Scale Measurement
Vin @ 50 MHz Vout @ 50
MHz
Calculated
Gain @ 50
MHz
Vin @ Max
Freq
Vout @ Max
Freq
Calculated
Gain @ Max
Freq Test Limit
is > -3dB
Channel 1
5mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div
500 mV/div
1V/div
Channel 2
5mV/div
10 mV/div
20 mV/div
50 mV/div
100 mV/div
200 mV/div

Table of Contents