EasyManua.ls Logo

Keysight Technologies Infiniium MXR254A

Keysight Technologies Infiniium MXR254A
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92 Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
6 Testing Performance
Channel 5
5 mV/div ±60 mV 3.1 mV 2.5 mV
10 mV/div ±120 mV 4.2 mV 3.0 mV
20 mV/div ±240 mV 6.4 mV 4.0 mV
50 mV/div ±600 mV 13.0 mV 7.0 mV
100 mV/div ±1.2 V 24.0 mV 12.0 mV
200 mV/div ±2.4 V 58.0 mV 22.0 mV
500 mV/div ±4.0 V 112.0 mV 52.0 mV
1 V/div ±4.0 V 162.0 mV 102.0 mV
Channel 6
5 mV/div ±60 mV 3.1 mV 2.5 mV
10 mV/div ±120 mV 4.2 mV 3.0 mV
20 mV/div ±240 mV 6.4 mV 4.0 mV
50 mV/div ±600 mV 13.0 mV 7.0 mV
100 mV/div ±1.2 V 24.0 mV 12.0 mV
200 mV/div ±2.4 V 58.0 mV 22.0 mV
500 mV/div ±4.0 V 112.0 mV 52.0 mV
1 V/div ±4.0 V 162.0 mV 102.0 mV
Channel 7
5 mV/div ±60 mV 3.1 mV 2.5 mV
10 mV/div ±120 mV 4.2 mV 3.0 mV
20 mV/div ±240 mV 6.4 mV 4.0 mV
50 mV/div ±600 mV 13.0 mV 7.0 mV
100 mV/div ±1.2 V 24.0 mV 12.0 mV
200 mV/div ±2.4 V 58.0 mV 22.0 mV
500 mV/div ±4.0 V 112.0 mV 52.0 mV
1 V/div ±4.0 V 162.0 mV 102.0 mV
Channel 8
5 mV/div ±60 mV 3.1 mV 2.5 mV
Table 7 Offset Accuracy Test (50 ) (continued)
Vertical Scale Ext.
Supply
Setting
V
DMM
+
V
Scop
e+
V
error
+
V
DMM
-
V
Scop
e-
V
error-
Offset
Accuracy
Limit (±)
V
zero-error
V
zero-error
Limit (±)

Table of Contents

Related product manuals