Page #24 background imageLoading...
Page #24 background image
2-4 Keysight M937xA PXIe Service Guide
Tests, Adjustments, and Troubleshooting
Before You Begin
2-
Equipment Critical Specifications Recommended
Model or Part
Number
Alternate Model
Number
Use
a
Test Instruments and Software
PNA
b
Published port match
specifications
N5241A/42A, or
N5221A/22A, or
E8363C/64C
None P
Compression test set
b
None specified U3070AK01 None P
Dynamic accuracy test set
b
None specified U3020AD01 None P
Frequency counter
b
Freq: 10 MHz to 20 GHz
Accuracy: ±0.5 ppm
53151A
Option 001
Any equivalent P, A
Signal generator
b
CW Freq: 1.185 GHz N5181A E8257D
Option 520
P
USB Thermocouple Power
Sensor
Freq: DC to 33 GHz
Range: –35 to +20 dBm
U8485A
Option 200
N8485A power sensor
(for Adjustments, also
requires N8482A for
frequencies below 10
MHz) with
N1912A/13A/14A
power meter
b
P, A
ECal module, 3.5 mm Freq: 300 kHz to 26.5 GHz N4691B 85052B P
Optional: Printer, Mouse,
Keyboard
N/A Any printer with Microsoft Windows 7 driver. --
Cables
BNC cable 50 ohm, length ³60 cm 8120-1839 Any equivalent A
3.5 mm test cable
(-m- to -f-), (Qty 3)
N/A 8121-2111
(60 cm), or
N4373-61604 (90
cm)
Any equivalent P
3.5 mm test cable
(-m- to -m-)
N/A 11500E Any equivalent A
GPIB controller cable N/A
M9036-31301
c
Any equivalent P, A
GPIB cable (Qty 2) N/A 10833A (1 m) Any P

Table of Contents

Other manuals for Keysight Technologies M9370A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies M9370A and is the answer not in the manual?

Keysight Technologies M9370A Specifications

General IconGeneral
BrandKeysight Technologies
ModelM9370A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals