7 Decrement the first digit until 0” is displayed.
You decrement or increment each digit independently. Neighboring
digits are not affected.
000.000,0 mVDC
8 Move the flashing cursor over to the units” location.
Notice that the units are flashing on the right side of the display.
000.000,0 mVDC
9 Increase the displayed number by a factor of 10.
Notice that the position of the decimal point changes and the displayed
number increases by a factor of 10.
0.000,000 VDC
10 Save the change and turn off the menu.
The multimeter beeps and displays a message to show that the change
is now in effect. You are then exited from the menu.
The zero offset calibration procedure determines new calibration
constants for each function and range. Separate calibration is required
for the front and rear terminals.
<
<
Auto/Man
ENTER
3
Chapter 3 Menu Tutorial
A Front-Panel Menu Tutorial
57