Error  Error Message 
605
606
701
702
703
704
705
706
707
708
709
720
721
722 [1]
723 [1]
725 [2]
730
731
732
733
734
735
736 [2]
740 [3]
741 [3]
742 [3]
743 [3]
744 [3]
745 [3]
746 [3]
747 [3]
748 [3]
Cannot calibrate rundown gain 
Rundown gain out of range 
Cal security disabled by jumper 
Cal secured 
Invalid secure code 
Secure code too long 
Cal aborted 
Cal value out of range 
Cal signal measurement out of range 
Cal signal frequency out of range 
No cal for this function or range 
Cal DCV offset out of range 
Cal DCI offset out of range 
Cal RES offset out of range 
Cal FRES offset out of range 
500V DC correction out of range 
Precharge DAC convergence failed 
A/D turnover correction out of range 
AC flatness DAC convergence failed 
AC low frequency convergence failed 
AC low frequency correction out of range 
AC rms converter noise correction out of range 
AC rms 100th scale linearity correction out of range
Cal checksum failed, secure state 
Cal checksum failed, string data 
Cal checksum failed, DCV corrections 
Cal checksum failed, DCI corrections 
Cal checksum failed, RES corrections 
Cal checksum failed, FRES corrections 
Cal checksum failed, AC corrections 
Cal checksum failed, GPIB address 
Cal checksum failed, internal data 
Calibration Error Messages 
[1] RES is the 2-wire ohms function. FRES is the 4-wire ohms function.
[2] Available starting with firmware Revision 2 (REV 02-01-01).
[3] This error may be generated if you turn off the power during a calibration.
Recalibration is required to clear the error. If the error persists
,
a hardware failure of U505 may have occurred.
Chapter 4  Calibration Procedures
Error Messages  
90