Zero Adjustment
(continued) 2 Select the shorted terminal set (front or rear terminal) with the
front/rear switch. Separate calibration constants are stored for the
front and rear input terminals.
3 Turn on the menu (
Shift < ) and then use < or > to
select
F: CAL MENU.
4 Use
to move down to the “commands” level and select
2. CALIBRATE.
5 Use
to move down to the “parameters” level and then set the
input value to 000.0000 mV.
6 Execute the command by pressing
Auto/Man .
7 Perform the Zero Offset Verification tests (see page 67) to check
zero calibration results .
8 Repeat steps 1 through 7 for the other input terminal set (front or
rear terminal).