MEASure:DIODe?
Sets all measurement parameters and trigger parameters to their default values for diode test meas-
urements and immediately triggers a measurement. The results are sent directly to the instrument's out-
put buffer.
Parameter Typical Return
(none) (none)
Configure, make, and read a default diode measurement:
MEAS:DIOD?
Typical Response:+1.32130000E-01
l
The range and resolution are fixed for diode tests: the range is 1 VDC (with a 1 mA current source out-
put).
l The voltage is displayed on the front panel if it is between 0 and 5.05 V. The instrument beeps when
the signal transitions into the range between 0.3 and 0.8 V (unless beeper is disabled). If the signal
exceeds 5.05 V, the front panel shows "OPEN", and the value returned from SCPI is 9.9E37.
l The FETCh?, READ?, and MEASure:DIODe? queries return the measured voltage, regardless of its
value.