EasyManua.ls Logo

Keysight 34921A - Page 57

Keysight 34921A
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56 Keysight 34921A-34925A Low Frequency Multiplexer Modules User’s Guide
Low Frequency Multiplexer Modules 34925A 40/80-Channel Optically-Isolated FET Multiplexer
Bank 2
For orientation, the D-sub connector
end of the module is facing you.
Bank 1
Bank 2
74
34
35
36 37
38
39 40
41
42
43 44 45
46
47 48 49
50
18
19 20
21 22
23
24
25 26
27 28
29 30 31 32
33
GND515271726364555675765960
NC NC NC NC
Reserved 61 62 53 54 73 65 66 57 58 77 78 69 70 Interlock 2
1 2 3 4 5 6 7 8 9 1011121314151617
Interlock 241 42 43 44 45 46
COM
2H
COM
2L
47 48 67 68 49 50 79 80
50-Pin D-Sub
Male Connector
Desc Pin Desc Pin Desc Pin Desc Pin Desc Pin
411513561197137COM2 H 7
422523662207238COM2 L8
433532163397323Interlock
2
17
444542264407424Interlock
2
33
455554165257543Reserved 18
466564266267644GND34
479572767117729No
Connect
47
48 10 58 28 68 12 78 30 No
Connect
48
49 13 59 45 69 31 79 15 No
Connect
49
50 14 60 46 70 32 80 16 No
Connect
50
As a safety feature, interlock 2 pins (17 and 33) on Bank 2
must be shorted to enable the Bank 2 Analog Bus relays to
close. The optional 34925T-002 (for 1-wire) terminal block
shorts these pins for you. This feature protects inadvertent
routing of high voltages from the Analog Buses to the
module.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com