32 Keysight E5063A Network Analyzer
Troubleshooting
Function Specific Troubleshooting
1-
Figure 1-16 Front Panel Test window 2/2
Step 4. To exit the front panel test, click the button.
If multiple keys fail to work, a problem in the A52 Front Panel I/F board or
A60 CPU Module is suspected. Also, check the flat cable between the A52
Front Panel I/F board and A60 CPU Module.
If only a specific key fails to work, check first if the key is subsided in the
panel.
If the rotary knob fails to work, check the A52 Front Panel I/F board
involving the RPG.
To Check the Touch Panel
Procedure
Using the LCD display panel, select or change the setting of a function in the
softkey menu and then, perform the same operation using the hardkeys.
If the touch panel does not work correctly when the hardkeys function
normally, a failure seems to be in the touch screen controller assembly or
touch-panel LCD assembly. (The touch panel is not replaceable
independently of the LCD.)
Check the cable between the touch screen controller and the serial
interface connector on the A60 CPU module.
If no problem is found in the above checks, a failure in the A60 CPU Module
is suspected.