Security Features & Document of Volatility 10
Instrument Memory & Volatility
MXG (Series A) Memory
MXG (Series A) Memory
The following tables describe each memory type used in the base instrument
and optional baseband generator.
Table 4-1 Base Instrument (All models and options)
Memory
Component,
Type and Size
Writable During Normal
Operation?
Data Retained When
Powered Off?
Purpose/Contents Data Input
Method
Location in
Instrument and
Remarks
Sanitization
Procedure
1. Main Memory
(DRAM)
32 MByte
Yes No Firmware operating
memory
Operating
system
CPU board, not
battery backed.
Turn off
instrument power.
2. Main Memory
(Flash)
8 MByte
Yes Yes Factory calibration/
configuration data
User file system
a
, which
includes flatness
calibration, instrument
states, and sweep lists
Firmware
upgrades and
user-saved
data
a
CPU board
(same chip as
firmware
memory, but
managed
separately)
Because this
memory chip
contains 8 MB of
user data
(described here)
and 8 MB of
firmware memory,
a full-chip erase is
not desirable.
User data areas
are selectively and
completely
sanitized by the
procedure “Erase
and Sanitize All
(Option 006 Only)”
on page 22.