Security Features & Document of Volatility 20
Memory Clearing, Sanitization and Removal Procedures
Erase All (Option 006 Only)
Erase All (Option 006 Only)
This function removes all user files, user flatness calibrations, user I/Q
calibrations, and resets all table editors with original factory values, ensuring
that user data and configurations are not accessible or viewable. The
instrument appears as if it is in its original factory state, however, the memory
is not sanitized. This action is relatively quick, typically taking less than one
minute (the exact time depends on the number of files).
Note that there is a similar but distinct function, as described below, that
deletes all user files but does not reset the table editors:
Model Key Sequence
MXG (Series A) File > More > Security > Erase All > Confirm Erase
ESG Utility > Memory Catalog > More > Security > Erase All > Confirm Erase
Model Key Sequence
MXG (Series A) File > Delete All Files
ESG Utility > Memory Catalog > More > Delete All Files