EasyManua.ls Logo

Keysight N1914A - Page 15

Keysight N1914A
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 15
Find us at www.keysight.com
E-Series CW Power Sensor Specifications (Continued)
Power linearity
Table 6. E4410 Series power linearity specification.
Power Temperature (25 ± 5 °C) Temperature (0 to 55 °C)
100 pW to 10 mW (70 to +10 dBm) ± 3% ± 7%
10 mW to 100 mW (+10 to +20 dBm) ± 4.5% ± 10%
+20
+10
+6
–20
–35
–70
± 4%
+3%
A
B
–70 –35 –20 –10 +10 +20
Power level used as reference (dBm)
Power level being measured (dBm)
Figure 1� Relative mode power measurement linearity with EPM Series power
meter/E-Series CW power sensor at 25 °C ± 5 °C (typical)�
The chart in Figure 1 shows the typical
uncertainty in making a relative power
measurement, using the same power
meter channel and the same power sensor
to obtain the reference and the measured
values� Example A illustrates a relative
gain (amplifier measurement)� Example
B illustrates a relative loss (insertion
loss measurement)� This chart assumes
negligible change in frequency and
mismatch occur when transitioning from
the power level used as the reference to
the power level being measured�
Example A
P = 10(P)/10 x 1 mW
P = 10 6/10 x 1 mW
P = 3�98 mW
3%x3.98mW=119.4μW
Example B
P = 10 (P)/10 x1 mW
P = 10 –35/10 x 1 mW
P = 316 nW
3% x 316 nW = 9�48 nW
where
P = power in Watts
and
(P) = power in dBm
Mechanical characteristic
Mechanical characteristics such as center
conductor protrusion and pin depth are
not performance specifications� They
are, however, important supplemental
characteristics related to electrical
performance� At no time should the pin
depth of the connector be protruding�

Other manuals for Keysight N1914A

Related product manuals